In carrying out the medical device registration process in Indonesia the applicant must fulfill administrative and technical documents requirements. However, during the registration process MoH still found that applicant does not provide complete documents which will cause on application rejection. MoH during the seminar on August 29th 2018, explain some important points that can be used by the applicant in preparing submission documents.
No |
Documents |
Important point |
A-3 |
Letter of Authorization |
|
A-7 |
Standard and declaration of conformity |
Attached are the standard reference, for example: |
B-11 |
Material |
Include a list of materials or components |
B-13 |
Manufacturing flowchart |
|
C-1 |
Finished Product Specifications |
|
C-3 |
Validation of the Sterilization Process |
|
C-4 |
Pre-clinical studies |
|
C-6 |
Result of the research on any device containing biologic material. |
|
C-8 |
Risk analysis of device |
Refer to ISO 14971: Medical devices - Application of risk management to medical devices or on GHTF / SG3 / N15R8 documents |
C-9 |
Result of Risk Analysis |
Refer to ISO 14971: Medical devices - Application of risk management to medical devices or on GHTF / SG3 / N15R8 documents |
C-10 |
Specification and/or requirement of raw material |
|
C-13 |
Results of analytical tests or results of clinical trials and safety of medical devices |
|
D-1 |
Labelling |
|
D-2 |
Symbol and Explanation |
List the codes or symbols that are listed on medical devices or packaging / labels |
D-3 |
Guidelines of usage, training material & guidance of installation and its maintenance |
Include a manual / insert package / |
D-4 |
Product code |
|
E-5 |
Standard Operating Procedure, record management and product complaints handling, etc. |
Complaint Handling Procedures and Forms and Recall Include: |
Other things to look for, including:
- - The product name must be consistent (in each document MUST be the SAME)
- - List of size / variant details do not forget to fill in the D-5 form
- - The product name listed on the CoA or Final QC Report must be in accordance with the product name registered
- - Product name writing includes: Brand + Generic Name
- - Example: BALLERINA Contact Lens