TAIWAN: Updates in the Product Registration Guidance for Artificial Intelligent/Machine Learning-Based Software as a Medical Device – April, 2022
                         2022-04-27
                      
                     
                   
                  Artificial Intelligent/Machine Learning based Software as Medical Device (SaMD) is now a hot topic in enhancing the welfare of humanity. To ensure the safety, performance, and quality of medical device software that utilizes artificial intelligence/machine learning technology, TFDA has established a guidance with reference to the regulations of the United States, Japan, and Korea and the International Medical Device Regulatory Forum (IMDRF). This guidance was initially announced in September 2020 and was further updated in August 2021.
The latest requirements are summarized in the table below. However, depending on the different characteristics of each SaMD, additional documents may still be required during the reviewing process. 
	
		
			| 
				Defining the software as: | 
				Artificial Intelligence, AIMachine Learning, MLDeep Learning | 
		
			| 
				Description of software functions: | Functional types, including- 
				Computer Assisted Detection, CADeComputer Aided Diagnosis, CADxComputer Aided Triage | 
		
			| Algorithm framework- 
				DesignTraining methodTesting principle and algorithm framework | 
		
			| 
				Data Limitation | 
				Shall include the detail of the training method, structure, process, data relevancy, and quality maintenance.Shall describe the training module database, in population, clinical meaning, output type, output method, and other additional clinical significancesThere are 3 kinds of data applied in SaMD, data for training, validation, or testing.Among all, data for testing must be strictly separated from data for training or validation to avoid bias in the validated results.
 | 
		
			| 
				User Environment and Data Management Safety | 
				Shall list out the limitations of operating environment and personnel.If the SaMD can be connected to the internet, has wireless transmission functions, or is a kind of mobile application (APP), shall refer to “Cybersecurity Guidelines for Medical Device manufacturers” and provide related documents. | 
		
			| 
				Functional Verification and Validation | Shall refer to “Software Medical Device Guidelines”, including- 
				Level of concernSoftware descriptionDevice hazard analysisSoftware requirements specificationsArchitecture design chartSoftware design specificationTraceability analysisSoftware development environment Verification and validation documentationRevision level historyUnresolved anomalies/ bugs or defects 
 Besides the above V&V documents, the output results shall comply with the intended use of the SaMD, including- 
				How the output result will be presentedThe limitation of the output resultSubsequent clinical suggestions regarding the output result | 
		
			| 
				Clinical Significance | Scientific evidence shall be provided to establish the applicability and suitability of the software specifications. It is recommended to include the below concept in the study protocol- 
				Intended useStudy objectivesPatient population, e.g. age, ethnicity, race…Number of clinicians and qualificationsDescription of the methodology used in gathering clinical informationDescription of the statistical methods used to analyze the dataStudy results | 
	
 
Reference: 
Product Registration Guidance for Artificial Intelligent/Machine Learning-based Software Medical Device